Operational Verification
6-4
Service Manual
6.5
Operational Checks
Ensure the ventilator control panel is adjusted to the settings shown in
Note:
Adjustments are made with the touch-turn-accept method.
Table 6-1.
Standard Test Settings
Control
Setting
Power
On
Patient type
Adult
Breath type
Volume control
Mode
A/CMV
FiO
2
0.60
Flow
30 L/min
Tidal volume
500 mL
Resp rate
10 b/min
Pressure limit
5
Pressure support
0
Exp.threshold
Auto
Open exh.
Off
Flow wave
Square
Slope rise
5
Volume target
Off
PEEP
0
Trig
Pressure 3.0 cmH
2
O
RR high alarm
110 b/min
MVE alarm
Low at 1 L/min, high at 50 L/min
Pause
Off
Paw alarm
Low at 10, high at 75
Leak comp.
Off
Circuit type
Test lung
Alarm loudness
1
Apnea alarm
60
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......