Operational Verification
6-12
Service Manual
6.5.13
PEEP/CPAP
1.
Set the ventilator to the standard test settings, except set PEEP to 3 cmH
2
O.
2.
Set the calibration analyzer to PEEP measurement.
Note:
Allow three to four breaths at each setting for pressure to stabilize.
3.
Verify that the value of PEEP displayed on
Numeric
screen, the pressure reading measured by the
calibration analyzer, the pressure bar graph on the control panel, and the
PEEP/CPAP
setting are within
±1 cmH
2
O of each other. Document the results on the record sheet.
4.
Repeat step 3 with a
PEEP
setting of 10 cmH
2
O.
5.
Set
Low Paw
alarm to 20 cmH
2
O.
6.
Repeat step 3 with a
PEEP
setting of 20 cmH
2
O.
7.
Set
PEEP
to 0 cmH
2
O and
Low Paw
alarm to 10 cmH
2
O.
6.5.14
Manual Inflation
1.
Set the ventilator to volume control SPONT.
2.
Press the
Manual Inflation
button.
3.
Verify that a manual breath is delivered while the button is pressed.
6.5.15
Inspiratory Tidal Volume
1.
Set the ventilator to volume control A/CMV.
2.
Set
Flow
to 30 L/min.
3.
Press the
Main Screen
button on the control panel.
4.
Touch the
Numeric
button on the screen.
5.
Set the calibration analyzer to volume measurement in BTPS with a 10 L/min threshold.
6.
Set
Tidal Volume
to 200 ml.
7.
Verify that the value of VTI displayed on the
Numeric
screen, the tidal volume reading measured by the
calibration analyzer, and the
Tidal Volume
setting are within ±10% of each other. Document the results
on the record sheet.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......