Operational Verification
6-8
Service Manual
6.5.6
Main Flow
1.
Remove the test lung from the patient circuit Y-piece.
2.
Place the calibration analyzer at the inhalation outlet.
3.
Set the calibration analyzer to flow measurement.
4.
Press the button under the
Tidal Volume
window.
5.
Use the touch-turn-accept method to adjust setting to 2.00 L.
6.
To choose
Flow
, press the
Select
button at the top of the
Flow/t Insp
window.
7.
Press the
Main Screen
button on control panel.
8.
To display the numeric data, touch the
Numeric
button on the screen.
Note:
When the Insp Flow reading displayed on the
Numeric
screen is higher than expected flow, it could be
caused by the peak flow delivered in the very beginning of the inspiratory phase. In this case, select
Flow Wave
on the
Extended Functions
screen, and then touch the
Freeze
button to read the actual flow
value. To adjust the scale of the waveform, touch the flow axis on the screen, turn the adjustment knob
to obtain a proper scale, and then touch the flow axis again to accept the change.
9.
Set
Flow
to 5 L/min.
10.
Verify that the value of Insp Flow displayed on the
Numeric
screen, the flow reading measured by the
calibration analyzer, and the flow setting are within ± 1L/min of each other. Document the results on the
record sheet.
Note:
The flow reading measured by the calibration analyzer mentioned in steps 10 through 16 is the actual
delivered measurement divided by 89%. This is because the ventilator automatically reduces the flow
delivery by 11% of the set flow to compensate for humidifier use (BTPS compensation).
11.
Set
Flow
to 10 L/min.
12.
Verify that the value of Insp Flow displayed on the
Numeric
screen, the flow reading measured by the
calibration analyzer, and the flow setting are within ±1 L/min of each other. Document the results on the
record sheet.
13.
Set
Flow
to 50 L/min.
14.
Verify that the value of Insp Flow displayed on the
Numeric
screen, the flow reading measured by the
calibration analyzer, and the flow setting are within ±4 L/min of each other. Document the results on the
record sheet.
15.
Set
Flow
to 80 L/min.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......