Operational Checks
Service Manual
6-11
Figure 6-4.
Pressure Control Test
6.5.11
Pressure Support
1.
Set the ventilator to pressure control SPONT.
2.
Set
Trig
to P (pressure) and adjust to 0.5 cmH
2
O.
3.
Set
Pressure Support
to 20 cmH
2
O.
4.
Squeeze the test lung to trigger a breath.
5.
Verify that the value of Ppeak displayed on the
Numeric
screen, the peak pressure measured by the
calibration analyzer, and the
Pressure Support
setting are within ±10% of each other. Document the
results on the record sheet.
6.
Set
Pressure Support
to 0 cmH
2
O.
7.
Verify no pressure-supported breaths are delivered.
6.5.12
Pressure Trigger Sensitivity
1.
Set
Pressure Support
to 10 cmH
2
O.
2.
Set
PEEP/CPAP
to 7 cmH
2
O.
3.
Set ventilator
Trig
to P (pressure) and adjust to 1.0 cmH
2
O.
4.
Manually squeeze and slowly release the test lung to create a dip in the pressure bar graph. Verify that
the dip equals the
P Trig
setting when a pressure support breath is triggered.
5.
Repeat step 4 with
Ptrig
setting of 3 cmH
2
O.
6.
Set
Pressure Support
to 0 cmH
2
O.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......