Component Removal and Replacement
4-18
Service Manual
Figure 4-13.
Pneumatics Panel
6.
Remove the yellow tube and the blue tube from the air and oxygen PSOL valves by disassembling the
luer lock connectors.
7.
While supporting the pneumatics panel, slide the panel away from the ventilator.
8.
To reinstall the panel assembly and components, perform the following steps:
a.
Supporting the pneumatics panel, align the air and oxygen flow sensors with the flow sensor
manifold block.
b.
Continuing to provide support, secure two screws on top for the pneumatics panel (once secured,
it is safe to release support).
c.
Reverse steps 2 through 6.
1
Screws (two of four securing PNL2102M)
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......