Calibration Procedures
5-6
Service Manual
5.
Remove the orange tubing from transducer XD107 on the analog PCB.
6.
Set respiratory rate to 128.
7.
To start the offset calibration, press the
Manual Inflation
button. The message “Start Calibration” is
displayed.
8.
Verify the
t Insp
setting is “0”. If not, perform the following steps:
a.
Press the button below the
Flow / t Insp
window. The display will blink.
b.
While the display is blinking, turn the adjustment knob to adjust the
t Insp
setting to “0”.
c.
To confirm the selection, press the
Accept
button.
9.
Press the button below the
Resp Rate
window.The display will blink.
10.
While the display is blinking, turn the adjustment knob to adjust the value.
11.
Press the
Accept
button to confirm the selection.
12.
After the
Accept
button is pressed, the value of Drive Pressure displayed on the touch screen changes.
•
As the setting of
Resp Rate
increases, the value of Drive Pressure will decrease after the
Accept
button
is pressed.
•
As the setting of
Resp Rate
decreases, the value of Drive Pressure will increase after the
Accept
button
is pressed.
Note:
A
Resp Rate
setting of “0” cannot be further decreased. Consequently, the value of Drive Pressure cannot
be increased. In this case, first increase the setting of
Resp Rate
to a high value and then decrease it.
Always press
Accept
when the adjustment of the
Resp Rate
setting is made.
13.
Repeat steps 8 to 12 to obtain a Drive Pressure reading of 3.00 ±0.05.
Table 5-2.
Zero Offset Calibration
Zero setting
Transducer ID
LCD display
t Insp
Pressure
Tolerance
Tube color
XD107
Drive Pressure
0
3.00
±0.05
Orange
XD106
Machine 2
2
0.00 cmH
2
O
±0.05
Green
XD105
Machine 1
4
0.00 cmH
2
O
±0.05
Clear
XD102
Air Gas Supply
6
0.0 psig
±0.1
Yellow
XD103
O
2
Gas Supply
8
0.0 psig
±0.1
Blue
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......