Calibration Procedures
5-2
Service Manual
•
Power source of 100–240 V AC, 50/60 Hz
•
High-pressure gauge for reading from 0–60 psig, or equivalent
•
Low-pressure gauge for reading from 0–200 cmH
2
O, or equivalent
•
Pressure syringe tool (TOL1952P) or equivalent
•
1/4 inch open-ended wrench
•
5/16 inch open-ended wrench
•
Medium-size flat head screwdriver
•
Low-pressure calibration tool (TOL2108A)
Caution:
The accuracy of all test equipment (electronic or pneumatic) used for verification or calibration
procedures should be certified annually by a testing laboratory before use.
5.3
Setup
Figure 5-1.
Newport™ e360 Ventilator Connected to Gas Sources
1
Air supply
2
Oxygen supply
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......