Software Upgrade Procedure
Service Manual
2-27
Figure 2-22.
Patient Circuit Setup
5.
Occlude the patient circuit using a cap (a filter and test lung are not required).
6.
Turn the ventilator back on.
7.
Wait until the
Ventilation Standby
screen appears.
8.
Follow the instructions that appear on screen, as shown in
1
Y-piece
2
Breathing circuit