Software Upgrade Procedure
Service Manual
2-29
Figure 2-25.
Circuit Check Screen (Step 2)
11.
When the circuit test is completed successfully, the message “Circuit Check PASSED” is displayed.
12.
If the circuit test failed, the message “Circuit Check FAILED” is displayed. Check the breathing circuit for
leaks, and repeat the circuit check. If the circuit check fails again, contact the Technical Service
Department for further assistance.
13.
Press the
Setup & Calibration
button on the control panel.
14.
Touch the
Technical
button on the screen (see
).
Figure 2-26.
Technical Screen
15.
Touch the
Regional Settings
16.
Touch the
Altitude
button on the screen. The “_ ft/_ m” setting will start flashing.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......