Procedures
Service Manual
5-15
3.
Occlude the patient circuit at the Y-piece using a cap (a filter and test lung are not required for this
calibration).
4.
To perform the circuit check, touch
Circuit Check
on the screen.
5.
Confirm that the circuit check passes.
6.
To start exhalation flow sensor calibration, perform the following steps:
a.
Touch
Sensors
on the screen.
b.
Press
Flow Sensor
.
Note:
You may need to perform step 6 several times to pass the exhalation flow sensor calibration.
7.
Enter diagnostics mode by performing the steps in section
8.
Set the ventilator to diagnostics mode “d9”.
9.
With the patient circuit occluded, press the
Manual Inflation
button to start the exhalation flow sensor
calibration.
10.
Wait for several minutes for the calibration to complete, and the message “Test Complete” to be
displayed.
Exhalation PSOL Valve: d4
1.
Enter diagnostics mode by following the procedure in section
2.
Touch “Start Diagnostics” on the screen.
3.
Press the button below the
Trig
window. The display will blink.
4.
While the display is blinking, turn the adjustment knob to change the setting from “d0” to “d4”.
5.
Press the
Accept
button to confirm the selection.
6.
With the patient circuit occluded, press the
Manual Inflation
button once. The message “Occlude the
circuit” is displayed.
7.
To start the calibration, press the
Manual Inflation
button again.
8.
Wait for several minutes for the calibration to complete, and the message “Test Complete” to be
displayed.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......