Electrical Safety Checks
Service Manual
6-3
4.
Connect regulated adjustable air and oxygen gas supplies to ventilator inlets on the back of the
ventilator.
5.
Set regulators to 50 psig.
6.
Use test form 10129540 to record the test data. Check the appropriate box or record the appropriate data
on the operational verification procedure record sheet after each test is performed.
6.4
Electrical Safety Checks
6.4.1
Ground Resistance Test
1.
Connect the AC power cord to the EST analyzer.
2.
Connect the ground lead of the EST analyzer to the labeled equipotential connection on the back of the
ventilator.
3.
Perform the ground resistance check, and record the results on the record sheet. To pass, the ventilator
ground resistance must be <0.1.
6.4.2
Current Leakage Test
1.
Turn the ventilator on.
2.
Perform the current leakage check, and record the results on the record sheet. To pass, the measured
current leakage must be <300 μA.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......