![background image](http://html1.mh-extra.com/html/covidien/newport-e360/newport-e360_service-manual_2673907126.webp)
Operational Verification
6-14
Service Manual
•
The audible alarm sounds
•
The internal battery indicator illuminates
•
At least three bars on the charge bar graph light up
3.
Reconnect the power cord to the AC power source.
Note:
To ensure that the internal battery remains functional, fully charge the battery at least every 3 months when the
ventilator is not in use.