Maintenance, Overhaul, and Software Upgrade
2-20
Service Manual
Figure 2-15.
Removing the Coin Battery
3.
Slide the new coin battery (PT00055561) into the retainer (BH1), ensuring that the positive side of the
battery is facing outward, away from the SBC board, as shown in
Figure 2-16.
Inserting a New Coin Battery
4.
Reconnect the internal battery positive (+) and negative (-) terminals.
5.
Reset the ventilator time and date:
a.
Turn on the ventilator.
b.
Press the Setup & Calibration key.
c.
Touch the Technical button.
d.
Enter the current date and time using the Hour, Minute, Month, Day, and Year buttons.