Maintenance Procedures
Service Manual
2-5
2.7
Maintenance Procedures
WARNING:
Disconnect electrical power, air, and oxygen sources before attempting any disassembly.
2.7.1
Inlet Water Trap Filters and O-Ring Assembly
Figure 2-1.
Inlet Water Trap Filter Replacement
Oxygen Filter
To replace the oxygen filter, perform the following steps:
1.
Unscrew the collection bowl from the inlet water trap.
2.
Unscrew the filter holder.
3.
Remove and replace the inlet filter and o-ring (JFK100P).