Component Removal and Replacement
4-6
Service Manual
4.
Locate and remove the following cables:
•
GR-CBL2101A from J2 on GR-PCB2100A
•
GR-CBL2118A from J13 on GR-PCB2100A
•
GR-CBL2148A from J8 on GR-PCB2100A
•
PT00095462 from J10 on GR-PCB2100A
•
10105498 air and oxygen PSOL valve cables from J705 (air) and J706 (oxygen) on GR-PCB2100A
•
GR-CBL2107A from J708 on GR-PCB2100A
•
GR-CBL2127A from J9 on GR-PCB2100A
•
GR-CBL2155A from CN4 and C9 SBC2105A.
5.
After all the cables from step 4 are disconnected, fully remove the front panel assembly and set on a
stable surface.
6.
Reinstall the front panel by reversing steps 2 through 5.
Note:
There is a small hole adjacent to J9 on GR-PCB2100A. The hole polarizes the connector.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......