Procedures
Service Manual
5-5
Figure 5-2.
Analog PCB
1.
Enter diagnostics mode by performing the steps in section
, “
.” “d0” is displayed in the
Trig
window on the control panel and “0” is displayed in the
Flow/t Insp
window.
2.
Press the button below the
Trig
window.The display will blink.
3.
While the display is blinking, turn the
Adjustment
knob to change the setting from “d0” to “d5”.
4.
To confirm the selection, press the
Accept
button.
Note:
Steps 5 to 15 are to calibrate zero pressure offset of the exhalation drive pressure transducer (XD107).
Note:
Reference
for tubing, transducer, and zero pressure offset values.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......