Calibration Procedures
5-14
Service Manual
Oxygen PSOL Valve: d2
Caution:
Do not attempt to perform oxygen PSOL valve calibration unless instructed by the Technical Service
Department or if one of the following conditions occurred:
•
Replacement or readjustment of oxygen inlet regulator
•
Replacement of oxygen PSOL valve
•
Replacement of analog PCB
•
Replacement of main PCB
Caution:
Do not attempt to perform oxygen PSOL valve calibration unless the oxygen gas supply is regulated to
50 ±2 psig and capable of supporting a minimum of 180 L/min of constant flow.
1.
Enter diagnostics mode by following the procedure in section
.”
2.
Touch “Start Diagnostics” on the screen.
3.
Press the button below the
Trig
window.The display will blink.
4.
While the display is blinking, turn the adjustment knob to change the setting from “d0” to “d2”.
5.
Press the
Accept
button to confirm the selection.
6.
Disconnect the patient circuit from the ventilator.
7.
To initiate oxygen PSOL valve calibration, press the
Manual Inflation
button.
•
The ventilator will go through a calibration maneuver that slowly increases and then decreases flow
twice.
•
The value of
TSI O2 Flow
displayed on screen will slowly change as the delivered flow changes.
8.
Wait several minutes for the calibration to complete, and the message “Test Completed” to be displayed.
Note:
If the message “Calibration Failed: Valve Leaks more than 50 cc” is displayed after the d2 is complete, the PSOL
valve needs to be replaced.
Exhalation Flow Sensor: d9
1.
Turn off the ventilator.
2.
Turn on the ventilator, and wait until the
Ventilation Standby
screen is displayed.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......