Section 7. Installation
7.1 Moisture Protection
When humidity tolerances are exceeded and condensation occurs, damage to
CR800 electronics can result. Effective humidity control is the responsibility of
the user.
Internal CR800 module moisture is controlled at the factory by sealing the module
with a packet of silica gel inside. The desiccant is replaced whenever the CR800
is repaired at Campbell Scientific. The module should not be opened by the user
except to replace the lithium coin cell providing back up power to the clock and
SRAM. Repeated disassembly of the CR800 will degrade the seal, leading to
potential moisture problems.
Adequate desiccant should be placed in the instrumentation enclosure to prevent
corrosion on the CR800 wiring panel.
7.2 Temperature Range
The CR800 is designed to operate reliably from -25 to +50°C (-40°C to +85°C,
optional) in non-condensing environments.
7.3 Enclosures
Illustrated in figure
Enclosure
(p. 82)
is a typical use of an enclosure, which is
available from Campbell Scientific. This style of enclosure is classified as
NEMA 4X (watertight, dust-tight, corrosion-resistant, indoor and outdoor use).
Enclosures have back plates to which are mounted the CR800 datalogger and
associated peripherals. Back plates are perforated on one-inch centers with a grid
of square holes that are lined as needed with anchoring nylon inserts. The CR800
base has mounting holes (some datalogger models may be shipped with rubber
inserts in these holes) through which small screws are inserted into the nylon
achors. Remove rubber inserts, if any, to access the mounting holes. Screws and
nylon anchors are included in an enclosure supply kit included with the enclosure.
Refer to
Enclosures
(p. 544)
for a list of available enclosures.
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......