Section 7. Installation
258
Table 50.
PRTCalc() Type-Code-6 Sensor
Standard ITS-90 SPRT, alpha = 0.003926 (Reference: Minco /
Instrunet)
Constant Coefficient
a 3.9848000E-03
d -2.3480000E-06
e 1.8226630E-05
f -1.1740000E-06
g 1.600
h -2.400
i 8.800
j 2.502
7.8.18.2 Measuring PT100s (100-Ohm PRTs)
PT100s (100-ohm PRTs) are readily available. The CR800 can measure PT100s
in several configurations, each with its own advantages.
7.8.18.2.1 Self-Heating and Resolution
PRT measurements present a dichotomy. Excitation voltage should be maximized
to maximize the measurement resolution. Conversely, excitation voltage should
be minimized to minimize self-heating of the PRT.
If the voltage drop across the PRT is
≤
25 mV, self-heating should be less than
0.001°C in still air. To maximize measurement resolution, optimize the excitation
voltage (Vx) such that the voltage drop spans, but does not exceed, the ±25-mV
input range.
7.8.18.2.2 PT100 in Four-Wire Half-Bridge
Example Shows:
•
How to measure a PRT in a four-wire half-bridge configuration
•
How to compensate for long leads
Advantages:
•
High accuracy with long leads
Example PRT Specifications:
•
Alpha = 0.00385 (PRT Type 1)
A four-wire half-bridge, measured with
BrHalf4W()
, is the best configuration for
accuracy in cases where the PRT is separated from bridge resistors by a lead
length having more than a few thousandths of an ohm resistance. In this example,
the measurement range is -10 to 40°C. The length of the cable from the CR800
and the bridge resistors to the PRT is 500 feet.
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......