Section 8. Operation
294
Other sensors, e.g., LVDTs (linear variable differential transformers), require an
ac excitation because they rely on inductive coupling to provide a signal. dc
excitation will provide no output.
CR800 bridge measurements can reverse excitation polarity to provide ac
excitation and avoid ion polarization.
Note
Sensors requiring ac excitation require techniques to minimize or eliminate
ground loops. See
Ground Looping in Ionic Measurements
(p. 91).
8.1.3.2 Accuracy of Ratiometric-Resistance Measurements
The ratiometric-accuracy specification for resistance measurements is:
±(0.04%
*
V1
+
Offset
),
‐
25°
to
50°
C,
where
V1
is
the
voltage
measurement
and
Offset
is
equal
to
one
of
the
following,
where
the
Basic
Resolution
is
the
resolution
of
a
single
A/D
(p.
425
)
conversion.
Note
that
excitation
reversal
reduces
offsets
by
a
factor
of
two:
•
Offset
= 1.5 x Basic Reso 1.0 µV if the measurement is made on a
differential input channel with input reversal
•
Offset
= 3 x Basic Reso 2.0 µV if the measurement is made on a
differential input channel without input reversal
•
Offset
= 3 x Basic Reso 3.0 µV if the measurement is of a single-
ended input channel
•
The
following
table
lists
basic
resolution
values.
Table 62.
Analog Input-Voltage Range and Basic
Resolution
Range (mV)
Basic Resolution (µV)
±
5000
±
2500
±
250
±
25
±
7.5
±
2.5
1333
667
66.7
6.7
2.0
0.67
Assumptions that support the ratiometric-accuracy specification include:
•
Excitation voltages less than 1000 mV are reversed during the excitation
phase of the measurement.
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......