Section 7. Installation
156
Scan
(100,mSec,0,0)
'Simulate measurement by exciting channel VX1/EX1
ExciteV
(Vx1,mV,0)
'Make the calibrated measurement
VoltSE
(RH,1,mV2500,6,1,0,250,Multiplier,Offset)
'Perform a calibration if CalMode = 1
FieldCal
(0,RH,1,Multiplier,Offset,CalMode,KnownRH,1,30)
'If there was a calibration, store it into a data table
CallTable
(CalHist)
NextScan
EndProg
7.8.1.5.2 Offset (Option 1)
Case: A sensor measures the salinity of water. Multiplier is known to be stable,
but sensor offset drifts and requires regular offset correction using a standard
solution. The following procedure offsets the measurement to obtain the
calibration report shown.
Table 23.
Calibration Report for Salinity Sensor
Parameter
Parameter at
Deployment
Parameter at 7-Day
Service
mV
output
1350
mV
1345
mV
KnownSalt
(standard
solution)
30
mg/l
30
mg/l
Multiplier
0.05
mg/l/mV
0.05
mg/l/mV
Offset
-37.50
mg/l
-37.23
mg/l
RH
reading
30
mg/l
30
mg/l
1. Send the program in CRBasic example
FieldCal Offset Demo Program
(p. 156)
to the CR800. An excitation channel has been programmed to simulate a
sensor output.
2. To simulate the salinity sensor in deployment-calibration conditions (30 mg/l
standard solution), place a jumper wire between channels
VX1
/
EX1
and
SE6
(
3L
). Set variable
mV
to
1350
. Set variable
KnownSalt
to
30
.
3. To simulate the deployment calibration, change the value in variable
CalMode
to
1
to start calibration. When
CalMode
increments to
6
, offset calibration is
complete. Calibrated offset will equal
-37.48
mg/l at this stage of the example.
4. To continue this example and simulate an offset-drift condition, change
variable
mV
to
1345
.
5. To simulate 7-day, service-calibration conditions (30 mg/l standard solution),
set variable
KnownSalt
to
30.0
. Change the value in variable
CalMode
to
1
to
start calibration. When
CalMode
increments to
6
, 7-day, service-offset
calibration is complete. Calibrated offset will equal
-37.23
mg/l.
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......