Appendix B. Status Table and Settings
511
Table 117.
Status-Table Fields and Descriptions
Fieldname Description
Variable
Type
Default
Range
Edit?
Info
Type
ProcessTime
Processing time (
μ
s) of the
last scan. Time is measured
from the end of the EndScan
instruction (after the
measurement event is set) to
the beginning of the EndScan
(before the wait for the
measurement event begins)
for the subsequent scan.
Calculated dynamically (on
the fly).
Integer
Status
MaxProcTime
Maximum time (
μ
s) required
to run through processing for
the current scan. This value
is reset when the scan exits.
Calculated dynamically (on
the fly).
Integer
Yes
Reset by
changing to 0
Status
BuffDepth
Shows the current Pipeline
Mode processing buffer
depth., which indicates how
far processing is currently
behind measurement.
MaxBuffDepth
Gives the maximum number
of buffers processing lagged
measurement.
LastSystemScan8
The last time the background
calibration executed.
Integer array
Status
LastSlowScan9
The last time
SlowSequence
scan(s) executed.
Integer array
Status
SystemProcTime8,12
The time (
μ
s) required to
process the background
calibration.
Integer array
Status
SlowProcTime9,12
The time (
μ
s) required to
process
SlowSequence
scan(s).
Integer array
Status
MaxSystemProcTime8,
13
The maximum time (
μ
s)
required to process the
background calibration.
Integer array
Status
MaxSlowProcTime9,13
The maximum time (
μ
s)
required to process
SlowSequence
scan(s).
Integer array
Status
PortStatus
Array of Boolean values
posting the state of control
ports. Values updated every
500 ms.
Boolean array of 8
False
True or False
Yes
Status
PortConfig
Array of strings explaining
the use of the associated
control port. Valid entries
are: Input, Output, SDM,
SDI-12, Tx, and Rx.
String array of 8
Input
Input or Output
Status
SW12Volts
Status of switched, 12-Vdc
port
Boolean False
True
or
False
Yes
Status
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......