Section 7. Installation
155
1. Send CRBasic example
FieldCal Zeroing Demonstration Program
(p. 155)
to the
CR800. An excitation channel has been programmed to simulate a sensor
output.
2. To place the simulated RH sensor in a simulated-calibration condition (in the
field it would be placed in a desiccated chamber), place a jumper wire between
channels
VX1
/
EX1
and
SE6
(
3L
). Set variable
mV
to
1000
. Set variable
KnownRH
to
0.0
.
3. To simulate a calibration, change the value in variable
CalMode
to
1
to start
calibration. When
CalMode
increments to
6
, zero calibration is complete.
Calibrated
Offset
will equal
-50
% at this stage of this example.
Figure 48: Zero (Option 0)
4. To continue this example and simulate a zero-drift condition, change variable
mV
to
1050
.
5. To simulate conditions for a 30-day, service-calibration, again with desiccated
chamber conditions, set variable
KnownRH
to
0.0
. Change the value in
variable CalMode to
1
to start calibration. When
CalMode
increments to
6
,
simulated 30-day, service zero calibration is complete. Calibrated
Offset
will
equal
-52.5
%.
CRBasic
Example
26.
FieldCal()
Zeroing
Demonstration
Program
'Jumper VX1/EX1 to SE6(3L) to simulate a sensor
Public
mV
'Excitation mV Output
Public
KnownRH
'Known Relative Humidity
Public
CalMode
'Calibration Trigger
Public
Multiplier
'Multiplier (Starts at .05 mg / liter / mV,
'does not change)
Public
Offset
'Offset (Starts at zero, not changed)
Public
RH
'Measured Relative Humidity
'Data Storage Output of Calibration Data -- stored whenever a calibration occurs
DataTable
(CalHist,NewFieldCal,200)
SampleFieldCal
EndTable
BeginProg
Multiplier = .05
Offset = 0
LoadFieldCal
(true)
'Load the CAL File, if possible
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......