Section 7. Installation
160
1. Send the program in CRBasic example
FieldCal Multiplier and Offset
Demonstration Program
(p. 160)
to the CR800.
2. To simulate the flow sensor, place a jumper wire between channels
VX1/EX1
and
SE6
(
3L
).
3. Simulate deployment-calibration conditions (output @ 30 l/s =
300 mV
, output
@ 10 l/s =
550 mV
) in two stages.
a. Set variable
SignalmV
to
300
. Set variable
KnownFlow
to
30.0
.
b. Start the deployment calibration by setting variable
CalMode
=
1
.
c.
When
CalMode
increments to
3
, set variable
SignalmV
to
550
. Set
variable
KnownFlow
to
10
.
d. Resume the deployment calibration by setting variable
CalMode
=
4
4. When variable
CalMode
increments to
6
, the deployment calibration is
complete. Calibrated multiplier is
-0.08
. Calibrated offset is
53.978
.
5. To continue this example, simulate a two-stage, 7-day service calibration
wherein both multiplier and offset drift (output @ 30 l/s = 285 mV, output @
10 l/s = 522 mV).
a. Set variable
SignalmV
to
285
. Set variable
KnownFlow
to
30.0
.
b. Start the 7-day, service calibration by setting variable
CalMode
=
1
.
c.
When
CalMode
increments to
3
, set variable
SignalmV
to
522
. Set
variable
KnownFlow
to
10
.
d. Resume the 7-day service calibration by setting variable
CalMode
=
4
6. When variable
CalMode
increments to
6
, the 7-day, service calibration is
complete. Calibrated multiplier is
-0.0842
. Calibrated offset is
53.896
.
CRBasic
Example
29.
FieldCal()
Multiplier
and
Offset
Demonstration
Program
'Jumper VX1/EX1 to SE6(3L) to simulate a sensor
Public
SignalmV
'Excitation mV output
Public
KnownFlow
'Known water flow
Public
CalMode
'Calibration trigger
Public
Multiplier
'Sensitivity
Public
Offset
'Offset (starts at zero, not changed)
Public
WaterFlow
'Water flow
'Data Storage Output of Calibration Data
─
stored whenever a calibration occurs
DataTable
(CalHist,NewFieldCal,200)
SampleFieldCal
EndTable
BeginProg
Multiplier = 1
Offset = 0
LoadFieldCal
(true)
'Load the CAL File, if possible
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......