Section 7. Installation
237
200
Table 37.
Parameters for Analog Burst Mode (601 to 2000 Hz)
CRBasic
Analog
Voltage
Input
Parameters
Description when in Burst Mode
Destination
A variable array dimensioned to store all measurements from a single channel.
For example, the command,
Dim
FastTemp(500)
dimensions array
FastTemp()
to store 500 measurements (one second of data at
500 Hz, one-half second of data at 1000 Hz, etc.)
The dimension can be any integer from
1
to
65535
.
Repetitions
The number of measurements to make on a single channel. This number
usually equals the number of elements dimensioned in the
Destination
array.
Valid arguments range from
1
to
65535
.
Voltage Range
The analog input voltage range to be used during measurements. No change
from standard measurement mode. Any valid voltage range can be used.
However, ranges appended with 'C' cause measurements to be slower than
other ranges.
Single-Ended
Channel
The single-ended analog input channel number preceded by a dash (
-
). Valid
arguments range from
-1
to
-6
.
Differential
Channel
The differential analog input channel number preceded by a dash (
-
). Valid
arguments range from
-1
to
-3
.
Measure Offset
No change from standard measurement mode.
False
allows for faster
measurements.
Measurements
per Excitation
Must equal the value entered in
Repetitions
Reverse Ex
No change from standard measurement mode. For fastest rate, set to
False.
Rev Diff
No change from standard measurement mode. For fastest rate, set to
False.
Settling Time
Sample interval in µs. This argument determines the measurement rate.
500 µs interval = 2000 Hz rate
750 µs interval = 1333.33 Hz rate
Integ
Ignored if set to an integer.
_50Hz
and
_60Hz
are valid for AC rejection but are
seldom used in burst applications.
Multiplier
No change from standard measurement mode. Enter the proper multiplier.
This is the slope of the linear equation that equates output voltage to the
measured phenomena. Any number greater or less than
0
is valid.
Offset
No change from standard measurement mode. Enter the proper offset. This is
the Y intercept of the linear equation that equates output voltage to the
measured phenomena.
7.8.13 String Operations
String operations are performed using CRBasic string functions, as listed in
String
Functions
(p. 480).
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...