Section 8. Operation
291
1
A/D (analog-to-digital) conversion time = 15 µs
2
Reps/No Reps -- If Reps > 1 (i.e., multiple measurements by a single instruction), no additional time
is required. If Reps = 1 in consecutive voltage instructions, add 15 µs per instruction.
8.1.3 Resistance Measurements
Many sensors detect phenomena by way of change in a resistive circuit.
Thermistors, strain gages, and position potentiometers are examples. Resistance
measurements are special-case voltage measurements. By supplying a precise,
known voltage to a resistive circuit, and then measuring the returning voltage,
resistance can be calculated.
Read More!
Available resistive bridge completion modules are listed in the
appendix
Signal Conditioners
(p. 539).
Five bridge measurement instructions are features of the CR800. Table
Resistive
Bridge Circuits -- Voltage Excitation
(p. 292) show circuits that are typically
measured with these instructions. In the diagrams, resistors labeled R
s
are
normally the sensors and those labeled R
f
are normally precision fixed (static)
resistors. Circuits other than those diagrammed can be measured, provided the
excitation and type of measurements are appropriate. CRBasic example
Four-wire
Full-bridge Measurement
(p. 293)
shows CR800 code for measuring and processing
four-wire full-bridge circuits.
All bridge measurements have the parameter
RevEx
, which has an option to make
one set of measurements with the excitation as programmed and another set of
measurements with the excitation polarity reversed. The offset error in the two
measurements due to thermal EMFs can then be accounted for in the processing
of the measurement instruction. The excitation channel maintains the excitation
voltage or current until the hold for the analog to digital conversion is completed.
When more than one measurement per sensor is necessary (four-wire half-bridge,
three-wire half-bridge, six-wire full-bridge), excitation is applied separately for
each measurement. For example, in the four-wire half-bridge, when the excitation
is reversed, the differential measurement of the voltage drop across the sensor is
made with the excitation at both polarities and then excitation is again applied and
reversed for the measurement of the voltage drop across the fixed resistor.
Calculating the resistance of a sensor that is one of the legs of a resistive bridge
requires additional processing following the bridge measurement instruction. The
table
Resistive-Bridge Circuits with Voltage Excitation
(p. 292)
lists the schematics
of bridge configurations and related resistance equations.
Содержание CR850
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 26: ...Table of Contents 26...
Страница 30: ...Section 2 Cautionary Statements 30...
Страница 32: ...Section 3 Initial Inspection 32...
Страница 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Страница 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Страница 78: ...Section 5 System Overview 78...
Страница 80: ...Section 6 CR800 Specifications 80...
Страница 95: ...Section 7 Installation 95 Figure 35 DevConfig OS download window Figure 36 Dialog box confirming OS download...
Страница 104: ...Section 7 Installation 104 Figure 43 Include File settings via DevConfig Figure 44 Include File settings via PakBusGraph...
Страница 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Страница 268: ...Section 7 Installation 268...
Страница 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Страница 385: ...Section 8 Operation 385 8 8 1 Data Display Figure 114 Displaying data with the keyboard display...
Страница 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Страница 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Страница 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Страница 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Страница 396: ...Section 8 Operation 396...
Страница 402: ...Section 9 Maintenance 402...
Страница 450: ...Section 11 Glossary 450...
Страница 504: ...Appendix A CRBasic Programming Instructions 504...
Страница 526: ...Appendix B Status Table and Settings 526...
Страница 530: ...Appendix C Serial Port Pinouts 530...
Страница 536: ...Appendix E FP2 Data Format 536...
Страница 550: ...Appendix F Other Campbell Scientific Products 550...
Страница 564: ...Index 564 WriteIO 464 Writing Program 108 X XML 448 XOR 473 Y Y intercept 141 142 Z Zero 155 166 Zero Basis 151...
Страница 565: ......