CPEC300/306/310 Closed-Path Eddy-Covariance Systems
54
TABLE 6-4. Variables from the
Run Station
Menu
Variable Name
Default
Description
Name of variable in Public
Table (in case no
CR1000KD available)
Pump Tmpr Ok
-
This is a display (read-only) variable
indicating whether the pump temperature
is within its operating range.
pump_tmpr_ok
Pump Tmpr
-
This is a display (read-only) variable
showing the temperature of the pump in
deg C.
pump_tmpr
Pump flow
8.0 L·min
-1
This is a display (read-only) variable
showing the volumentric air flow to the
pump. If using vortex intake, this
includes both flow through sample cell
and through vortex bypass.
pump_flow
System Diag
-
System diagnostic word. If set to 0, there
are no error conditions detected. For more
details on system diagnostic word, see
system_diag
Auto Z/S on
FALSE
Set to TRUE to initiate an automatic zero
and CO
2
span of the gas analyzer. The
system will return
Auto Z/S on
to
FALSE once the zero/span is initiated,
and the system will return to EC field
measurements upon completion of the
zero/span. This variable is ommitted if
the system is a CPEC300 or CPEC306
(i.e., no valve module).
prfm_auto_zero_span_flg
Pump Off
FALSE
For CPEC300 and CPEC306 systems,
select
TRUE
to disable the pump. Set to
FALSE
to enable the pump. For
CPEC310 systems, the options are
PMP_OFF
to disable the pump,
FLD_MEA
to re-enable pump and
resume normal EC measurements, or
AUTO_ZS
to initiate an automatic
zero/span cycle.
sample_pump_off_flg
EC155_PW_on
-
This is a read-only variable indicating
whether the EC155 is powered on.
EC155_actual_pwr_on
EC155 Messg
-
This is a read-only string describing the
current status of the EC155.
message
EC155 Off
FALSE
Set to TRUE to power down the EC155.
Set to FALSE to power on EC155.
ec155_pwr_off_flg
6.4 Data Retrieval
The program stores a very limited amount of data to the internal CPU of the
datalogger, so a microSD Flash card should be used with the CR6. TABLE
shows the number of days of data a 2 GB, 8 GB, and 16 GB card will typically
hold before the memory is full and data starts to be overwritten. In cases where
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...