Appendix H. Output Variables
H-2
TABLE H-1. Values stored in table ts_data
Compile
Switch
Variable
Name
Units
Description
pump_flow
LPM
Sample pump speed, converted to volumetric flow rate
V
valve_number
Valve number (0 to 6); determines which zero/span valve is
selected
V
valve_flow
LPM
zero/span flow
D
CO2_signal
Relative strength of the infrared signal in the CO
2
absorption
band
D
H2O_signal
Relative strength of the infrared signal in the H
2
O absorption
band
D
source_tmpr
ºC
Temperature measured inside the source housing
D
pump_control
Pump speed control variable; can be from 0 (off) to 1 (full speed)
D
pump_press
kPa
Pressure measured at the sample pump inlet
D
pump_tmpr
ºC
Temperature of the pump
D
ControlBits
Status of heaters and fans (see Appendix
(p. I-1)
, for
details)
D, V
SequenceStep
Status of the zero/span sequence
D, V
valve_tmpr
ºC
Temperature of the valve module
D, V
valveControl
Zero/span flow control valve; can be from 0 (closed) to 1 (open)
D, V, S
scrub_press
kPa
Pressure measured at the scrub pump outlet
D, V, S
scrub_tmpr
ºC
Temperature of the scrub module
D
BattVolt
V
Supply voltage; measured inside datalogger after diode voltage
drop
D
process_time
µs
Time to process the scan (a copy of ProcessTime from the Status
table)
D
buff_depth
Number of scans in the pipeline buffer that have not yet been
processed (a copy of BuffDepth from the Status table)
flux
The online flux calculations are stored in table
flux
. A record is written to this
table at the end of every output interval (30 min). Space is allocated on the
memory card for 9,600 records (200 days at one record per 30 min). The CPU
has storage allocated for 480 records (10 days).
Most of the values stored in table
flux
are always stored, but some are
dependent on the settings for program constants. See Section
(p. 36)
, for details on setting program constants.
The first six values (
Hs
through
u_star
) are the basic online flux calculations.
The next five values (
Ux_Lag_Avg
through
sonic_samples
) are averages of
the sonic anemometer outputs. These are averages of the values that have been
lagged to align the sonic and IRGA data. They are filtered to include only those
data for which there are no sonic diagnostic flags.
The next six values (
CO2_Lag_Avg
through
irga_samples
) are the
corresponding lagged averages of the EC155 data.
The next two values (
pump_flow_Lag_avg
and
pump_OK_samples
) are the
corresponding lagged averages for the pump flow.
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...