Appendix A. EasyFlux DL CR6CP Process Flow
A-2
Send sonic data to covariance tables to be included for 5 min and
averaging interval covariances
Parse out gas diagnostic data, and filter bad gas data from being
included in statistical data
⇩
Calculate climate and gas variables (e.g., e, rho_d, rho_a, Td,
CO2_mixratio, H2O_mixratio, RH)
⇩
Store gas data into multiple datasets or temporary tables that each
have a different lag relative to sonic data (to be used later in cross
correlation; lags from
–
MAX_LAG TO +MAX_LAG are used; MAX_LAG
default is 2). For each dataset with a particular lag, recalculate cell and
ambient climate variables.
⇩
If using FW, store raw data in multiple datasets, each dataset with a
different lag applied to FW data relative to sonic data (to be used later
in cross correlation; lags from -MAX_LAG to +MAX_LAG are used;
MAX_LAG default is 2)
⇩
Control pump speed to achieve pressure set point
⇩
Measure rain gauge
⇩
Record time series measurement diagnostics in the
Diagnostic
output
table
⇩
Ingest new raw sonic and gas measurements from the EC100
Every SLOWSEQUENCE_SCAN_INTERVAL (default 5 s)
Measure CR6 panel temp
⇩
Measure biomet and energy balance (slow response) sensors
⇩
Calculate albedo
⇩