CPEC300/306/310 Closed-Path Eddy-Covariance Systems
17
a full-featured CRBasic program editor.
PC400
does not support combined
communication options (for example, phone-to-RF), PakBus routing, or
scheduled data collection.
LoggerNet
:
LoggerNet
is a full-featured software package that supports
programming, communication, and data collection and display.
LoggerNet
consists of a server application and several client applications integrated into a
single product. This package is recommended for applications that require
telecommunications support, scheduled data retrieval, or for large datalogger
networks.
4.10 Replacement Parts
Vortex Filter:
For EC155 analyzers with a vortex intake, the bypass line from
the vortex has a filter that will become clogged over time (typically many
months) and requires replacement. The filter consists of a 25 µm particulate
filter with ¼-in Swagelok nuts on either side as shown in FIGURE
Replace the filter when the signal strength has dropped to 0.8 or less.
FIGURE 4-20. Vortex filter for EC155 intake
Sonic Wicks:
A sonic wicks spares kit is used to replace the wicks on the
CSAT3A. The kit includes three top wicks, three bottom wicks, an installation
tool, and adhesive (see FIGURE
).
FIGURE 4-21. Sonic wick spares kit
Silica Desiccant Bags:
Silica desiccant bags (FIGURE
) are used to
desiccate the CPEC300-series system enclosure and should be periodically
replaced. These can be purchased as a single, four-unit pack or as a quantity of
20.
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...