Appendix H. Output Variables
H-5
TABLE H-2. Values stored in table flux
Compile
Switch
Variable Name
Units
Description
source_tmpr_Avg
ºC
Average source temperature
pump_ON_Avg
Fraction of time the pump was on
pump_flow_Avg
LPM
Average pump flow
pump_control_Avg
Average pump control value
pump_press_Avg
kPa
Average pump pressure
diag_CPEC_Avg
Average of diag_cpec
fast_samples
Number of samples included in the averages for
values measured in the main scan
buff_depth_Max
Maximum buffer depth during the output interval
process_time_Avg
µsec
Average processing time
PumpTmprOK_Avg
Fraction of time the pump temperature was within
its operating range
pump_tmpr_Avg
ºC
Average pump temperature
pump_heat_Avg
Fraction of time the pump heater was on
pump_fan_Avg
Fraction of time the pump fan was on
panel_tmpr_Avg
ºC
Average datalogger wiring panel temperature
BattVolt_Avg
V
Average supply voltage
BattVoltOK_Avg
Fraction of time the system was not shut down
because the supply temperature had dropped below
the limit
slow_samples
Number of samples included in the average for
values Measured in the slow scan
V
ValveTmprOK_Avg
Fraction of time the valve module temperature was
within its operating range
V
valve_tmpr_Avg
ºC
Average valve module temperature
V
valve_heat_Avg
Fraction of time the valve module heater was on
V
valve_fan_Avg
Fraction of time the valve module fan was on
V, S
scrub_press_Avg
kPa
average pressure at the scrub pump outlet
V, S
ScrubTmprOK_Avg
Fraction of time the scrub module temperature was
within its operating range
V, S
scrub_tmpr_Avg
ºC
Average scrub module temperature
V, S
scrub_heat_Avg
Fraction of time the scrub module heater was on
V, S
scrub_fan_Avg
Fraction of time the scrub module fan was on
LWS
leaf_wetness_Avg
Average leaf wetness sensor reading
LWS
intake_heater_Avg
V
Average intake heater voltage
zero_span
The
zero_span
table contains data from the automated zero/span sequences.
Records are written to this table when the calibration sequence is run (either
automatically initiated based on time, or manually initiated). Each zero/span
sequence will put multiple records in the
zero_span
table, corresponding to the
steps in the sequence. The actual number of records will depend on which
options are chosen. If a step is skipped, no record will be written to table for
that step.
The records in the
zero_span
table are distinguished by the first two values:
Sequence_Step (1 to 11) and V
alveNumber
(0 if no valve is selected, or 1 to
6). Space is allocated on the card for 33,000 records (approximately 200 days
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...