CPEC300/306/310 Closed-Path Eddy-Covariance Systems
58
TABLE 6-7. Data Fields in the Time_Series Data Output Table
Data Field Name
Units
Description
Data Field
Included
sampling_regime
Binary
number
For CPEC310 this identifies the current sampling
regime. Bits 0 through 3 correspond to the current
sampling site, and bit 4 is the omit flag. For
computing fluxes, filter all values not equal to 1.
(1 corresponds to site 1 which is sampling
ambient air with no omit flag set.) See Appendix
for more information on sampling regimes and
sites.
If system is a
CPEC310
FW
deg C
Air temperature measured by fine-wire
thermocouple
If FW05, FW1, or
FW3 is used
TABLE 6-8. Data Fields in the Diagnostic Output Table
Data Field Name
Description
Data Field
Included
system_diag
System diagnostic word. If 0, no system errors
detected. See Appendix
system diagnostic.
Always
sonc_er
Sonic error flag (set to TRUE or -1 if any sonic
diagnostic flag detected)
Always
irga_er
EC155 error flag (set to TRUE or -1 if any irga
diagnostic flag detected)
Always
pump_tmpr_er
Pump temperature error flag
Always
pump_flow_er
Pump flow error flag
Always
valv_tmpr_er
Valve temperature error flag
If CPEC310 with
valve module
valv_flow_er
Valve flow error flag
If CPEC310
scrb_tmpr_er
Scrub temperature error flag
If scrub module
used
diag_sonic
CSAT3A diagnostic word. If 0, no error flags set.
See EC155 user manual for more details.
Always
sonic_amp_l_f
Amplitude low diagnostic flag
Always
sonic_amp_h_f
Amplitude high diagnostic flag
Always
sonic_sig_lck_f
Signal lock diagnostic flag
Always
sonic_del_T_f_f
Delta Temp diagnostic flag
Always
sonic_aq_sig_f
Acquiring signal diagnostic flag
Always
sonic_cal_err_f
Calibration error diagnostic flag
Always
diag_irga
EC155 diagnostic word. If 0, no error flags set.
See EC155 user manual more details.
Always
irga_bad_data_f
Any gas analyzer diagnostic flag is set
Always
irga_gen_fault_f
General system fault diagnostic flag
Always
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...