![Campbell CPEC300 Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/campbell/cpec300/cpec300_product-manual_489189113.webp)
Appendix C. Wiring the CR6 and Optional Energy Balance Sensors
C-5
TABLE C-6. Default Wiring for Radiation Measurements Option 2
Sensor
Quantity
Wire Description
Color
CDM-A116
Terminal
NR01 4-Way
Radiometer
0 or 1
Pyranometer Up Signal
Red (cbl 1)
CDM Diff 1H
Pyranometer Up
Reference
Blue
1/
(cbl 1)
CDM Diff 1L
1/
Pyranometer Down Signal
White (cbl 1)
CDM Diff 2H
Pyranometer Down
Reference
Green/Black
1/
(cbl 1)
CDM Diff 2L
1/
Pyrgeometer Up Signal
Brown/Grey or Orange (cbl 1)
CDM Diff 3H
Pyrgeometer Up
Reference
Yellow
1/
(cbl 1)
CDM Diff 3L
1/
Pyrgeometer Down Signal Purple or Pink/Brown (cbl 1)
CDM Diff 4H
Pyrgeometer Down
Reference
Grey/Green
1/
(cbl 1)
CDM Diff 4L
1/
PT100 Signal
White/Yellow (cbl 2)
CDM Diff 5H
PT100 Reference
Green (cbl 2)
CDM Diff 5L
Current Excite
Red (cbl 2)
CDM X2
Current Return
Blue (cbl 2)
CDM AG
⏚
Shields
Clear
CDM AG
⏚
CNR4 4-
Way
Radiometer
0 or 1
Pyranometer Up Signal
Red
CDM Diff 1H
Pyranometer Up
Reference
Blue
1/
CDM Diff 1L
1/
Pyranometer Down Signal
White
CDM Diff 2H
Pyranometer Down
Reference
Black
1/
CDM Diff 2L
1/
Pyrgeometer Up Signal
Grey
CDM Diff 3H
Pyrgeometer Up
Reference
Yellow
1/
CDM Diff 3L
1/
Pyrgeometer Down Signal
Brown
CDM Diff 4H
Pyrgeometer Down
Reference
Green
1/
CDM Diff 4L
1/
Thermistor Signal
White
CDM Diff 5H
Thermistor V Excite
Red
CDM X2
Thermistor Reference
Black
CDM AG
⏚
Shields
Clear
CDM AG
⏚
1/
Jumper to
⏚
with user-supplied wires
Содержание CPEC300