Appendix L. CPEC300 Series Pump Replacement
L-5
8.
Remove pump connector from the pump electronics (FIGURE
).
FIGURE L-8. Location of pump connector in CPEC300 pump
electronics
9.
Gently lift the pump assembly from foam, leaving the tubes attached.
Turn it over and remove the two self-tapping #6 screws that attach the
pump to the metal box, as shown in FIGURE
FIGURE L-9. Self-tapping screws attaching pump to metal box