Appendix H. Output Variables
H-10
TABLE H-4. Values stored in table message_log
Compile
Switch
Variable
Name
Units
Description
V
valve_number
Valve number (0 to 6);
determines which zero/span
valve is selected
V
press_offset
kPa
Sample cell pressure measured
with no flow; used to infer valve
flow
V
valveTctl_ON
Boolean flag: True if valve
module temperature control is
enabled
V
valve_flow
LPM
Zero/span flow
V
valve_tmpr
ºC
Temperature of the valve
module
V, S
scrub_tmpr
ºC
Temperature of the scrub
module
V, S
scrub_press
kPa
Pressure measured at the scrub
pump outlet
buff_depth
Number of scans in the pipeline
buffer that have not yet been
processed (a copy of BuffDepth
from the Status table)
config_history
The user may change the configuration of the CPEC300 program at any time,
Configure the EasyFlux DL Program
(p. 34)
. All the
configuration variables and compile switches are written to the
config_history
table at startup, when a configuration variable is changed, and once a day at 15
minutes past midnight. This provides a historical record to be used as a
troubleshooting aid.
The values stored in table
config_history
are shown in TABLE
TABLE H-5. Values stored in table config_history
Variable Name
Units
Description
TIMESTAMP
TS
Date and time the record was written
RECORD
RN
record number
BATT_LOWLIMIT
V
CPEC300 will power down if supply voltage falls below this limit
BATT_DEADBAND
V
CPEC300 will power up if supply voltage rises this much above
the lower limit
SONIC_AZIMUTH
degrees
Angle between true north and the direction the CSAT3A is
pointing
BANDWIDTH
Hz
EC100 digital filter bandwidth
AMB_PRESS_NOMINAL kPa
Nominal ambient pressure for CPEC300 elevation
PUMP_SETPT
LPM
Flow setpoint for sample pump speed control
CAL_INTERVAL
min
Time between automatic zero/span sequences
CAL_TIMEOFFSET
min
Time offset for starting automatic zero/span sequences
ZERO_TIME
s
Time to flow zero gas
SPAN_TIME
s
Time to flow CO
2
span gas
CAL_FLOW_SETPT
LPM
Flow setpoint for zero/span
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...