CPEC300/306/310 Closed-Path Eddy-Covariance Systems
31
5.3.2 EC Sensor Cables
Ensure the EC100 is not powered. Connect the EC155 gas analyzer head,
EC155 sample cell, and CSAT3A sonic anemometer head to the EC100
electronics (FIGURE
).
FIGURE 5-8. EC155 and CSAT3A electrical connections; mounting
hardware and tubing not shown)
CPEC300-series instruments that are ordered with a CR6, are pre-
wired with the appropriate EC100 wiring. For users that need to
wire the system, follow the next sections to wire the connection
between the EC100 and the CPEC300-series enclosure.
Wire the SDM communications cable (CABLE4CBL-L) between the EC100
and the CPEC300-series enclosure as shown in FIGURE
shows the color scheme of the SDM wires.
TABLE 5-1. SDM Wiring
Description
Wire Color
EC100
CR6
SDM Data
Green
SDM-C1
SDM-C1
SDM Clock
White
SDM-C2
SDM-C2
SDM Enable
Red/Brown
SDM-C3
SDM-C3
Digital Ground
Black
Ground
Ground
Shield
Clear
Ground
Ground
NOTE
EC155
Analyzer Cable
CSAT3A Cable
EC155 Sample-cell Cable