TABLE 6-9. Data Field in the EC100_Config_Notes Output Table
Data Field Name
Units
Description
Data Field Included
press_source
–
A parameter indicating the sensor used by
EC100 for ambient pressure (0 for EC100
Basic Barometer, 1 for user/custom
barometer, 2 for EC100 Enhanced
Barometer)
Always
diff_press_select
–
A parameter indicating whether to report
ambient pressure (open-path system) or
differential pressure (closed-path system). It
should be set to 2, which will auto-select.
Always
sample_cell_press_type
–
A parameter indicating whether the sample
cell pressure sensor is differential or
absolute. Set to 0 for EC155 gas head serial
numbers less than 2000, and set to 1 for
serial numbers 2000 and greater.
Always
tmpr_source
–
A parameter indicating the sensor used by
EC100 for ambient temperature (0 for
EC100 Temperature Probe, no other values
valid)
Always
CO2_zero_coeff
–
CO
2
zero coefficient set from last CO
2
zero
Always
CO2_span_coeff
–
CO
2
span coefficient set from last CO
2
span
Always
CO2_span_mixra
µmol·mol
-1
CO
2
mixing ratio of span gas
Always
H2O_zero_coeff
–
H
2
O zero coefficient set from last H
2
O zero
Always
H2O_span_coeff
–
H
2
O span coefficient set from last H
2
O span
Always
H2O_span_T_DP
deg C
Dew point temperature of span gas
Always
EC155_pwr_off
-
A parameter indicating whether the EC155
gas head is in sleep or power-down mode.
Set to 0 for power on; set to 1 for power off.
Always
CR6_Volts
V
Battery voltage as measured at the CR6
battery input terminal
Always
Heater_volts
V
Heater Control Setting (–1 for disabled, –2
for auto control)
Always
Shadow_corr
–
Application of transducer shadowing
correction (0 for off, 1 for on)
Always
TABLE 6-10. Data Fields in the Flux_AmeriFluxFormat Output Table
Data Field Name
Units
Description
Data Field
Included
TIMESTAMP_START YYYYMMDDHHMM Start time of the averaging period
Always
TIMESTAMP_END
YYYYMMDDHHMM End time of the averaging period
Always
CO2
µmol·mol
-1
CO
2
flux after corrections
Always
CO2_SIGMA
µmol·mol
-1
Standard deviation of CO
2
Always
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...