Next, compare
scrub_tmpr
to the operating range (5 to 50 °C). The scrub
module will be disabled if it is too cold. The scrub module has a heater that
turns on if
scrub_tmpr
falls below 7 °C. If the scrub module temperature is
too low, check the operation of the heater which is controlled by public
variable
scrub_heat_ON
. For diagnosing a problem using data saved in the
output table
ts_data
, the state of fans and heaters is encoded into variable
ControlBits
to conserve memory space. See Appendix
(p. I-1)
.
This value is saved only if saving all diagnostics. Its corresponding variable
scrub_heat_Avg
is saved in the averaged output tables (
Flux
and
Zero_Span
). If the heater is on and the scrub module is too cold, check the
ambient temperature. The CPEC300 is rated for temperatures from
−
30 to
50 °C. If the system is started in cold weather after being turned off for several
hours, it may take up to 20 minutes to warm up to operating temperature.
The scrub module will also be disabled if it is too warm (above 50 °C). The
scrub module has a fan that turns on if
scrub_tmpr
rises above 45 °C. The fan
will stay on until the scrub temperature falls below 40 °C. If the scrub module
temperature is too high, check the operation of the fan which is controlled by
public variable
scrub_fan_ON
. For diagnosing a problem using data saved in
the output table
ts_data
, the state of fans and heaters is encoded into variable
ControlBits
to conserve memory space. See Appendix
(p. I-1)
.
This value is saved only if saving all diagnostics. Its corresponding variable,
scrub_fan_Avg,
is saved in the averaged output tables (
Flux
and
Zero_Span
).
If the fan is on and the scrub module temperature is too high, check the
ambient temperature. The CPEC300 is rated for ambient temperatures between
–30 to 50 °C.
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...