TABLE 6-12. Data fields in the Flux_Notes Output Table
Data Field Name
Units
Description
Data Field
Included
T_panel_CDMA_x
Deg C
Average panel temperature of the
CDM-A116, where x at the end of the
name is an index from 1 to 4,
representing the each of the thermistors
under the terminal strips of the CDM-
A116
If a CDM-
A116 is used
batt_volt
volt
Average battery voltage supplying
power to the datalogger
Always
slowsequence_Tot
count
Number of slow sequences during the
averaging interval (for example, the
number of times biomet and energy
balance sensors were measured)
Always
process_time
ms
Average processing time for each scan
Always
process_time_SIGMA
ms
Standard deviation of scan time
Always
process_time_Max
ms
Maximum processing time for a scan
Always
process_time_Min
ms
Minimum processing time for a scan
Always
buff_depth
number
Average number of records stored in the
buffer
Always
buff_depth_Max
number
Maximum number of records stored in
the buffer
Always
TABLE 6-13. Data Fields in the Zero_Span_Notes Table
Data Field Name
Units
Description
Data Field Included
CO2_gain
Adimensional
CO
2
gain factor
[CO2_span_gas/(CO2_span –
CO2_zero)]
Always
H2O_gain
Adimensional
H
2
O gain factor
[H2O_span_gas/(H2O_span –
H2O_zero)]
Always
CO2_zero
µmol
∙
mol
-1
CO
2
mixing ratio during zero (i.e.,
offset)
Always
H2O_zero
mmol
∙
mol
-1
H
2
O mixing ratio during zero (i.e.,
offset)
Always
cell_tmpr_zero
° C
Sample cell temperature during zero
Always
cell_press_zero
kPa
Sample cell pressure during zero
Always
scrub_press_zero
kPa
Scrub module gauge pressure during
zero
If scrub module used
CO2_sig_strgth_zero
Adimensional CO
2
signal strength during zero
Always
H2O_sig_strgth_zero
Adimensional H
2
O signal strength during zero
Always
diff_press_zero
kPa
Pressure difference between sample
cell and ambient during zero
Always
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...