CPEC300/306/310 Closed-Path Eddy-Covariance Systems
56
constant
ONE_FULL_TABLE
from
FALSE
to
TRUE
(p. 36)
, for details on changing constants).
TABLE 6-6. Data Output Tables
Table Name
Description
Recording Interval
Memory on
CR6 CPU
Memory on
SD Card
Time_Series
Time series data
(aligned to
account for
electronic
delays)
SCAN_INTERVAL
(default 100 ms)
Auto-Allocate
(typically less than
1 hour)
Time_Series is
broken up into 1
day files (see
Diagnostic
Reports most
recent diagnostic
flags from gas
analyzer and
sonic
anemometer
SCAN_INTERVAL
(default 100 ms)
1 record (most
recent scan)
0 records
EC100_Config_Notes
Reports settings
for the gas
analyzer and
sonic
anemometer
Whenever settings
are changed or
system is power
cycled
128 records
10*DAY_FLUX_CRD
records
(default 300
records)
Flux_AmeriFluxFormat
Processed flux
and statistical
data following
reporting
conventions and
order of
AmeriFlux
OUTPUT_INTERVAL
(default 30 minutes)
NMBR_DAY_CPU
(default 7 days)
Broken up into
30-day files; see
calculate number
of files
Flux_CSFormat
Processed flux
and statistical
data
OUTPUT_INTERVAL
(default 30 minutes)
NMBR_DAY_CPU
(default 7 days)
Broken up into
30-day files; see
calculate number
of files
Flux_Notes
Intermediate
variables, station
constants, and
correction
variables used to
generate flux
results
OUTPUT_INTERVAL
(default 30 minutes)
NMBR_DAY_CPU
(default 7 days)
Broken up into
30-day files; see
calculate number
of files
ZeroSpan_Check_Notes
Summary of
field calibration
data at the time
of an automatic
zero or span
When an automatic
zero or span is
performed
ONLINCAL_SZE_CPU
(default 7 days)
DAY_FLUX_CRD
(default 30 days)
System _Operatn_Notes
Records any
change in system
status
When there is a
change in system
status
500 records
10*DAY_FLUX_CRD
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...