Appendix L. CPEC300 Series Pump Replacement
L-7
FIGURE L-12. Pump side with inlet and outlet tubing connected
2.
Reattach the pump to metal box with two self-tapping #6 screws on
the back of the pump module electronics plate (FIGURE
Be careful not to pinch the fan wires under the pump and do not
overtighten screws.
3.
Reattach the pump connector to the pump electronics (FIGURE
4.
Place the pump assembly back into the foam on the shell cover. Make
sure the fan is in the hole in the end and the small tab on the metal box
is seated in the slit in the foam.
FIGURE L-13. Proper positioning of CPEC300 in shell cover
5.
Hold the pump assembly securely to the shell cover while replacing
the shell cover to the shell bottom. Make sure the fan does not slide
back out of its hole in the foam. Fasten the shell cover in place with
the six #4 screws (see FIGURE
6.
Reconnect the tubing to the inlet and outlet of the filter assembly (see
step 2 of removal and refer to FIGURE
NOTE