give a description of all data fields found in each
data output table and when each data field is included in the table.
Prior to coordinate rotations, the orthogonal wind components
from the sonic anemometer are denoted as U
x
, U
y
, and U
z
.
Following coordinate rotations, the common denotation of u, v,
and w is used, respectively.
Variables with _R denote that the value was computed after
coordinate rotations were done. Variables with a _F denote that
the value was calculated after frequency corrections were applied.
Similarly, _SND and _WPL refer to variables that have had the
SND correction or the WPL correction applied, respectively.
TABLE 6-7. Data Fields in the Time_Series Data Output Table
Data Field Name
Units
Description
Data Field
Included
Ux
m·s
-1
Wind speed along sonic x-axis
Always
Uy
m·s
-1
Wind speed along sonic y-axis
Always
Uz
m·s
-1
Wind speed along sonic z-axis
Always
T_SONIC
deg C
Sonic temperature
Always
diag_sonic
–
Raw sonic diagnostic value (0 indicates no
diagnostic flags set)
Always
CO2
µmol·mol
-1
CO
2
dry molar mixing ratio
Always
H2O
mmol·mol
-1
H
2
O dry molar mixing ratio
Always
diag_irga
–
Raw gas analyzer diagnostic value (0 indicates no
diagnostic flags set)
Always
TA_1_1_1
deg C
Air temperature calculated from sonic
temperature and humidity
Always
T_cell
deg C
Sample cell temperature
Always
PA_cell
kPa
Air pressure inside sample cell
Always
CO2_sig_strgth
–
CO
2
signal strength
Always
H2O_sig_strgth
–
H
2
O signal strength
Always
PA_diff
kPa
Differential pressure (the difference between
sample cell pressure and ambient pressure)
Always
PA
kPa
Ambient pressure
Always
pump_flow
L·min
-1
Volumetric air flow to pump; if using vortex
intake, this includes sample flow and vortex
bypass flow
Always
NOTE
NOTE
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...