Appendix C. Wiring the CR6 and Optional Energy Balance Sensors
C-3
TABLE C-3. Default Wiring for Fine-Wire Thermocouple
Sensor
Quantity
Wire Description
Color
CDM-A116 Terminal
FW05, FW1, or
FW3
0 or 1
Signal
Purple
CDM Diff 6H
Signal Reference
Red
CDM Diff 6L
Shield
Clear
CDM AG
⏚
C.1.5 Temperature and Relative Humidity Probe
The
EasyFlux DL CR6CP
can support from zero to one temperature and
relative humidity probe. The default wiring for the HMP155A or EE181 is
shown in TABLE
shows wiring for the HMP155A and EE181
temperature and humidity probes. Alternatively, an older model
probe such as the HMP45C or HC2S3 may be used, but wiring
for these models is not shown here. Instead, their wiring may be
found in the respective product manuals found at
. In any case, be careful to note the colors
of the wires and jumper wire configuration of the probe being
used. TABLE
shows wire colors for the HMP155A and
EE181, where the colors for the EE181 are noted by italic text in
TABLE C-4. Default Wiring for Temperature and Relative Humidity Probe
Sensor
Quantity
Wire Description
Color
CDM-A116
Terminal
HMP155A/
EE181
0 or 1
Temp Signal
Yellow/
Yellow
CDM Diff 13H
Temp Signal Reference
White/
Jumper to AG
⏚
CDM Diff 13L
1/
RH Signal
Blue/
Blue
CDM Diff 14H
RH Signal Reference
White/
Jumper to AG
⏚
CDM Diff 14L
1/
Shield
Clear/
Clear
AG
⏚
Power
Red/
Red
+12 V
Power Ground
Black/
Black
G
1/
For the HMP155A, a jumper wire should be added to connect CDM Diff 13L to CDM Diff 14L. For the EE181, jumper
wires should go from CDM Diff 13L and CDM Diff 14L to AG
⏚
.
C.1.6 Radiation Measurements Option 1
There are two options for making radiation measurements with
EasyFlux DL CR6CP
. The program can support any combination of the four
sensors described in TABLE
. Alternatively, it can support one of the two
types of four-way radiometers described in TABLE
default wiring for Option 1. TABLE
shows the details of the default wiring
for Option 2.
NOTE
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...