6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
112
Result units
Create a new unit with
NEW
and store it under a
new, unique name.
Select a base unit for the new unit calculation
Selectable are:
g N/L
,
% N
,
g Pr/L
or
% Pr
(N = nitrogen content, Pr = protein content)
Enter a correction factor and/or correction offset
if necessary and select the preferred number of
decimal places.
(The results are calculated following the linear equa-
tion
y= a·x+b
where
a
is given by the correction
factor and
b
by the correction offset.)
Dosage volume in status view
Set the increment in
[mL/click]
for the direct
dosing of H
2
O, NaOH and H
3
BO
3
within the status
view. (Per click on the corresponding dose button,
the specified amount in mL will be dosed.)