6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
70
Calibration pH electrode
It is recommended to calibrate the electrode
every day before starting sample determinations.
The electrode should be treated according to the
recommendation described in the electrode supple-
mentary sheet.
We recommend replacing the electrode, if it does
not fulfill the following criteria at 25 °C room temper-
ature anymore:
Slope
95 – 105 %
Zero point
pH 6.4 – 7.6
(For pH electrodes other than the ones supplied by
BUCHI, additional criteria might be important.)
NOTE
It is recommended to use buffer solutions pH 4.00
and 7.00. (For a 3 point calibration in addition the
buffer solution for pH 9.21 is recommended.)
Discard buffer solutions after usage. Work with
fresh solutions every day.