6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
117
6 9 2 Utilities
There are three different utilities contained in the
device software:
∙ The
Database management
– backup and
restore of the complete device data (settings,
methods, resources, results etc.)
∙ The
Lab timer
– create a countdown timer
with acoustic alarm.
∙ The
Demo mode
– use the device in demo
mode (without chemicals).
Database management
The database backup can be stored on a USB
device or on a network directory or on both in
parallel.
∙ Select
YES
for a „Use USB backup device“
and/or
YES
for a „Use network backup share“
∙ Specify the path to the USB and/or network
directory.
For the network share the IP address, name, domain
and User name and password have to be specified.
NOTE
For the network settings configuration the help of
an IT specialist of the company may be required!
Lab timer
∙ Create or delete a timer and define the count-
down time.
∙
Start
or
Stop
the selected timer.
NOTE
After the countdown time has elapsed, an acoustic
alarm is triggered. To stop this alarm sound press
the upper
STOP
button (Do not press the red
STOP button – this will stop all device processes!).
The volume of the alarm can be adapted within the
„Display and sound“ settings. (See chapter “6.9.1
Settings“)