7 Maintenance
K-375 / K-376 / K-377
Operation Manual, Version E
134
∙ Pull the sealing cap out of the
sampler arm from below.
∙ When inserting the new sealing cap
from below take care to its orienta-
tion – the two fittings on top are not
arranged symmetrically. When the
sealing cap is inserted properly, both
fittings have to reach through the two
corresponding holes in the sampler
arm.
∙ The red hose belongs to the fitting
with the red marking, the white hose
to the one with the white marking.
∙ Retain the sealing cap from below
while reattaching the two hoses to
prevent it from slipping out again.
∙ Push a new dip tube as far as it will
go into the sealing cap from below.
∙ Move the sampler arm to the zero
position. (System Preparation
Sampler functions
Move to zero
position)
∙ Reinstall the sample rack.
NOTE
Depending on sample througput and instrument care, an exchange of the sealing cap should be consid-
ered after around 2000 distillations. If steam leaks, the sealing cap needs to be exchanged immediately.
7 4 3
Replacing the splash protector
Replace the glass splash protector after approximately 3000 distillations, at the latest after 5000 distilla-
tions. The plastic splash protector needs to be replaced after around 8000 distillations.
For the replacement of the splash protector the open end spanner (11058252) and the tool SVL 22
(11057779) are needed. Both are part of the standard delivery.
!
WARNING
Serious chemical burns by corrosives.
∙ Observe all data sheets of the used chemicals.
∙ Handle corrosives in well ventilated environments only.
∙ Always wear protective goggles.
∙ Always wear protective gloves.
∙ Always wear protective clothes.
∙ Do not use damaged glassware.