6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
79
For samples of type
Control Blank
,
press “Name” and enter a name for the control
blank result.
Press “Method” and select the method to be used
for the determination of the control blank from the
list of the available methods.
Press “Group” and select a result group for the
storage of the result from the list of the available
result groups. (It is also possible to create a new
result group using the
New Group
button.)
6 6 3
Sample Lists
With the sample list functionality it is possible to predefine a complete list of samples to be deter-
mined one by one without a sampler. Each sample list can be filled with any number of predefined
samples. If all samples of a sample list are selected at once for the determination, they will be deter-
mined in the same order they were added to the list. It is also possible to determine the samples in a
different order by selecting individual samples from the list.
The type (Blank, Sample, Reference substance or Control blank), and name of each sample can
be chosen freely. The same applies for the used method and the result group for the storage of the
result.
For samples in addition the weight and the protein factor have to be specified. If a compatible balance
is connected to the instrument, the weight of each sample can be automatically taken over from this
balance.
NOTE
For each new sample the entries of the previously entered item are used as default values (the default
value for the name depends on the type of the sample – in this case the name of the last sample of
the same type is taken as default). All default values can be overwritten.
To start a sample list determination you have to enter the list and to select the samples to be deter-
mined. To select a complete list simply check the check box in front of the first sample and check and
hold the check mark in front of the last sample in the list or use
SELECT ALL
. As a result all samples
in between will also be checked. (This will also work for deselecting a larger number of samples.) To
exclude samples from the determination uncheck the check mark in front of the sample in question.
Once a sample has been determined (with or without a valid result) it is deleted from the list and the
next sample in the list becomes sample number one (the next sample to be determined). After all
samples of a list have been processed, the empty list remains on the device (it can either be refilled
with samples for the next determinations or deleted manually).