8 Troubleshooting
K-375 / K-376 / K-377
Operation Manual, Version E
152
14’602
Titrator stopped, maximal volume
reached
Make sure the electrode is in good working
conditions, the correct titration solution is used
and that no air bubbles are in titrator hoses.
14’603
Titrator stopped, stop endpoint
reached
Check the titrator and if enough titration solu-
tion is available
14’604
Titrator stopped, stop Potential
reached
Check the titrator and if enough titration solu-
tion is available
14’605
Titrator stopped, stop time reached
Check the titrator and if enough titration solu-
tion is available
15’001
No sampler connected
Switch on sampler and check the connection
cable
15’002
Sampler: Target not achieved
Try again or call service
15’003
Sampler: Connection lost
Check the connection cable
15’101
Sampler: Shield open
Close shield
15’102
Sampler: Crash detected
Try again or call service
15’103
Sampler: Tube not found
Put sample tube in place or call service
15’104
Sampler: Tube not released
Remove sample tube or call service
15’105
Sampler: Error 5, reserve (does not
exist)
Call service
15’106
Sampler: Error 6, reserve (does not
exist)
Call service
15’107
Sampler: Shield not locked
Call service
15’108
Sampler: Error reference position
Call service
15’109
Sampler zeroadjustment not possible
because x or y deviation is larger then
3mm or sampler was not in reference
position before adjustment start
Try again or call service
15’110
Sampler: Position error X-axis
Call service
15’111
Sampler: Position error Y-axis
Call service
15’112
Sampler: Position error Z-axis down
Call service
15’113
Sampler: Position error Z-axis up
Call service
15’114
Sampler: Error during EEPROM
writing. Adjustment value not saved
Call service
17’001
Steam generator overtemperature
Switch distillation unit off and on or call service
17’002
Water level not reached
Check steam generator water supply or call
service
18’001
Stirrer out of order
Check stirrer cable or exchange stirrer