4 Description of function
K-375 / K-376 / K-377
Operation Manual, Version E
43
5
Putting into operation
This chapter describes how the device is installed and gives instructions on initial startup.
NOTE
Inspect the device for damages during unpacking. If necessary, prepare a status report immediately
to inform the postal company, railway company or transportation company.
Keep the original packaging for future transportation.
!
CAUTION
Heavy weight, avoid overexertion.
∙ Due to the heavy weight of the devices at least two people are required for
taking the KjelMaster K-375 or the KjelSamplers K-376 out of their corre-
sponding packaging. Watch your fingers when putting the device down.
∙ For the K-377 at least three people are required for taking the device out of the
corresponding packaging. Watch your fingers when putting the device down.
5.1
Installation site
The device must be set-up on a stable, clean and level surface.
For safety reasons the distance between the back of the device and the wall or to another object
must be at least 30 cm. No containers, chemicals or other objects must be located behind the instru-
ment.
The KjelSampler K-376 or K-377 is set-up
on the left side
of the KjelMaster K-375 with a space of
approximate 10 cm. Make sure that the back of the KjelSampler is not in contact with anything, e.g.
hoses, etc.
All devices must be set up in such a way that the main switches and the mains plugs are easily
accessible at all times.
NOTICE
Risk of device damage.
∙ The sampler arm of the KjelSampler K-376 / K-377 must have enough space in
height for movement.
!
CAUTION
Heavy weight, avoid overexertion.
∙ At least two people are required for carrying the KjelSampler K-376 or the Kjel-
Master K-375 due to the heavy weight of the devices. Watch your fingers when
putting the devices down.
∙ At least three people are required for carrying the KjelSampler K-377 due to the
heavy weight of the device. Watch your fingers when putting the device down.